Analytical system |
Fully automated, random-access chemistry system with STAT capability |
|
Analytical principle |
Spectrophotometry and potentiometry |
|
Analytical types |
Endpoint, rate, fixed point and indirect ISE |
|
Analytical methods |
Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE |
|
Simultaneously processed analytes |
60 photometric tests + 3 ISE, 120 pre-programmed onboard tests |
|
Throughput |
400 photometric tests/hour; maximum of 800 with ISE |
|
ISE throughput |
200 samples/hr.Maximum test/hr 600 if ISE only |
|
Sample types |
Serum, plasma, urine, other |
|
Sample feeder |
Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 80 samples; continuous loading |
|
Sample tubes |
Primary and secondary tubes; diameter from 11.5 to 16 mm; height from 55 to 102 mm, nestled micro sample cups |
|
STAT samples |
Up to 22 positions for STAT samples via cooled turntable. Easy to operate interrupts between tests Auto repeat run capability |
|
Sample volume |
1–25 μl in 0.1 μl steps (1–25 μl for repeats) |
|
Reagent capacity |
76 positions for (R1+R2, detergent position). Holds 15, 30 and 60ml bottles |
|
Reagent storage |
Refrigerated (4 °C–12 °C |
|
Reagent volume |
R1: 10–250 μl; R2: 10–250 μl; (in 1 μl increment) |
|
Total reaction volume |
90–350 μl |
|
Reaction cuvette |
Permanant glass cuvettes |
|
Reaction time |
Up to 8 minutes, 37.5 seconds |
|
Reaction incubation |
37 °C, dry bath |
|
Photometry system |
Direct assay through the reaction cuvette (0 – 3.0 OD) mono and bichromatic measurements possible |
|
Wavelength |
13 different wavelengths between 340 – 800 nm |
|
Calibration |
Auto calibration, cooled calibrator positions; master calibration established by two-dimensional barcode |
|
Quality control |
Auto QC, cooled QC positions |
|
Rerun |
1. Auto rerun and manual rerun available 2. Automatic sample increase, decrease or normal repeat 3. Reflex testing |
|
Sample Pre Dilution |
Dilution ratio: Fixed to 3, 5, 10, 15, 20, 25, 50, 76 or 100 |
|
Test requisition |
Individual and profile test requisition via on-line, mouse, function keys and touch screen |
|
Sample integrity |
Lipaemia, Haemolysis and Icterus analysis. Sample clot detection and probe crash protection |
|
Online |
RS232C. Full uni- and bidirectional communication possible |
|
Software |
Windows XP® |
|
Data Storage |
Sample number 100,000 samples. Reaction Data 200,000 tests |
|
Dimensions (W x H x D) mm |
ANL: 1,450 x 1,205 x 760 |
|
Power supply |
100–240v; 60 Hz/ < 3.5 kVA |